Centauri Field Upgrade Instructions

Field Upgrade of the firmware is a very simple process. Centauri has the necessary infrastructure for performing field upgrade of the embedded software.


Download the latest Centauri firmware upgrade package:
  • The upgrade package is contained as a folder within a ZIP file. Extract from the ZIP file the folder (called 'centauri_upgrade') and copy this folder to main directory (the root) of a USB flash drive.
    Note: The USB flash drive must be up to 32G and formatted as FAT32
  • Insert the USB flash drive into the USB port that is found on the upper panel of the meter (see Figure 1).

    StarLite
    Figure 1: Upper panel of Centauri Meter with USB input highlighted
  • Insert a pin in the pinhole (see Figure 2) to access the micro-switch.



    Figure 2 Right side of Centauri Meter with pinhole highlighted
  • Turn on the Centauri while simultaneously pressing through the pinhole for a few seconds.
  • The upgrade procedure starts, with a progress bar displayed on the screen.
  • When the procedure is completed, the Centauri restarts with the new firmware version.

Disclaimer:

  • The new Centauri firmware is based on a new Android version so once upgraded there is no way back other than loading the instrument with production firmware.
  • Upgrade from older version may take over 5 and a half minutes.
  • The logs stored in the instrument will be deleted (Save them in a USB disk drive before the instrument is upgraded)