Dan Little, Technical Director, Laser Training Institute, Professional Medical Education Association, INC
Lasers used in the medical field fall under stricter safety regulations than other laser usages. Meeting ANSI compliances are critical to the continued legal operation of all medical and aesthetic facilities. Laser output powers and energies are to be checked on a semi-annual basis according to FDA Regulations and are supported by ANSI recommendations which state regular scheduled intervals. In our service school we exclusively use Ophir-Spiricon Laser Measurement Instrumentation. We present a graphically enhanced presentation on measurement technologies and the many, varying, critical parameters that are faced with not only each different type of laser but design differences between manufacturers.
Ophir offers a plethora of measurement heads that can satisfy any need. Needs, such as, beam diameters, multiple wavelengths, nanosecond pulsewidths and very low or high power and energy levels. As if these aren’t enough variables to deal with, space confinement can be a problem. Ophir also manufactures small heads to get into the tiny places. For example, I have a 20C-SH head that fits nicely between optical components inside our KTP lasers. The same head sits securely by itself on top of the fiber launch output on our argon laser. This physical compatibility works great when we are peaking output power with the OC mirror adjustments. Capturing nanosecond, mJ pulses from an Ophthalmic Q-Switched, Nd:YAG laser is always a good time. As a side note, always collect the reading on the prefocus side the impact. The defocus side will have a diminished reading because of losses in the photoacoustic shockwave at the focal point.
IPLs (Intense Pulsed Light) though not laser, are used in the aesthetic market because of their large spot sizes. Lasers, known for their ability to be focused into very tiny spot sizes, become less desirable when performing light-based hair removal on legs and backs. Large spot sizes can treat the areas in a fraction of the time. However, when checking output energies, trying to jam that large spot into a traditional laser head aperture is messy. Ophir, recently introduced a head for IPLs with a huge 65mm aperture. Whether the laser operates in continuous wave mode (CW), pulsed or Q-switched, Ophir-Spiricon is our first choice for testing and/or calibration measurement. Most common medical and aesthetic lasers are: CO2 (10,600 nm), Er:YAG (2,940 nm), Ho:YAG (2,100 nm), Nd:YAG (1064nm), Diode Lasers (950/810/etc), Alexandrite (755 nm), Ruby (694 nm), KTP (532 nm), Argon (515/488 nm), Dye (577 nm), Pulsed Dye (585/595 nm/etc), Xenon Chloride (308 nm), Argon Fluoride (193 nm).
CO2 (10,600 nm) - Because of its very long and far-IR wavelength, the CO2 beam has a very shallow absorption depth and a great affinity for water and almost everything including glass and fiber optics. Though the beam must be delivered via mirrors mounted in an articulated arm, the CO2 makes a great surgical “light” scalpel and ablator. When used with a scanner or pattern generator, cosmetic skin resurfacing is easily achieved. The treatment can be fractional or totally ablative.
Er:YAG (2,940 nm) – Similar to the CO2 but having a greater affinity for water, it creates a shallower surface ablation in skin resurfacing. This makes the treatment less traumatic but also less effective.
Ho:YAG (2,100 nm) – Also an affinity for water but able to transmit through fiber optics making it a unique surgical tool for Urology and Orthopedic surgery.
- up to ~ 5J max (most biomeds use a cw detector head and just use 10hz)
Nd:YAG (1064 nm) – This wavelength, just outside the visible spectrum, has the longest absorption depth. Because of this attribute, it is used in surgery (via fiber optics) as a deep tissue coagulator. With special, light absorbing fiber tips, it is used for cutting. This laser is also very popular in the aesthetic field for a multiple of treatments. Depending upon specific balances of pulse width, spot size and power levels, it is used for hair removal, small leg vein ablation, skin rejuvenation, tattoo removal and more.cw up to 100 W
Diode Lasers (950 nm /810 nm/etc) – Critically depending upon their varying wavelengths and delivery settings, diode lasers are used in dentistry, dermatology, ophthalmology, aesthetics and surgery.
Alexandrite (755 nm) – A popular laser for hair removal on lighter skin types only (the shorter wavelength than the Nd:YAG is absorbed more by melanin and can cause hyper/hypo pigmentation problems in darker skin types).
Ruby (694 nm) – The first laser invented is used for tattoo removal (Q-switched only) and hair removal on very light skin (skin type I)
KTP (532 nm) – The KTP laser, also called a “Frequency-Doubled YAG”, is actually an Nd:YAG laser (or diode) with an add-on KTP crystal. The 1064 nm wavelength is transmitted through the KTP crystal which then, by resonance induces a frequency an “octave” above the excitation frequency. In other words, 1064 nm (when doubled in frequency) becomes 532 nm. This is a green wavelength, half the wavelength of the 1064 nm. Green wavelengths are absorbed by their complementary color pigment (see your color wheel chart). This makes them very useful for ablating reds and browns as in vascular and pigmented lesions.
- quasi cw, 40w max (one “new” KTP for urology is up to 180 watts but this is uncommon still)
Argon (515/488 nm) – The green 515nm wavelength of the argon is generally used more than the blue 488nm in medicine. Like the 532nm green of the KTP, it is used for the same treatments. However, the argon 515 nm wavelength is much more preferred in Ophthalmology for retinal photocoagulation than the 532 nm because of its greater affinity for hemoglobin. This translates into less discomfort for the patient.
- cw, most up to 5W. A few up to 20W
Dye (577 nm) = This laser, capable of operating in a continuous wave modality, is generally used as an optical pumping source to excite the dye used to create a red 632 nm beam. This dye laser is mainly used in Photodynamic Therapy (PDT) destroy cancerous tumors.
Pulsed Dye (585 nm /595 nm/etc) – The yellow and orange wavelengths of the pulsed dye are used in dermatology and ophthalmology to target their complementary color pigments, which are mainly reds (vascular). The difference between using the yellow versus the green wavelengths for vascular, is that the yellow has no affinity for melanin. This results in less risk in patient injury and better tolerated by the patient.
Xenon Chloride (308 nm) – Delivered via fiber optics, the XeCl laser is used primarily for Angioplasty and general vessel clearing.
Argon Fluoride (193 nm) – This is only laser that performs “Lasik” eye surgery. Contrary to popular belief, the mechanism used here to re-shape the cornea, is NOT heat or ablation. This unique and very short wavelength borders on tissue damaging UV. The 193nm wavelength itself does the work. With not much power, the 193 nm light collapses the 3D carbon molecules into a 2D configuration wherever the light irradiates the tissue. This mechanism of laser treatment is called Photo-disassociation. With computer control of this “soft” ablation of tissue, a cornea can be reshaped to a less positive, convex lens, thus changing the dioptor of the eye optics. This corrects the focal length of the eye optics to meet the retina as eyeglasses would.